Abstract
The optimal diffusion weighting (DW) factor, b, for use in diffusion tensor imaging (DTI) studies remains uncertain. In this study, the geometric relations of DW quantities are examined, in particular, the effects of Rician noise in the measured magnetic resonance signal. This geometric analysis is used to make theoretical predictions for selecting a b value to reduce the influence of noise. It is shown that the optimal b value for DTI studies in healthy human parenchyma is approximately b=1200 s mm-2, with a simple relation given as well for a given expected apparent diffusion coefficient. Monte-Carlo simulations on sets of realistic DTI measures are then performed, verifying the optimal DW for minimizing estimate errors. The effects of noise on various DTI parameters such as anisotropy indices (fractional anisotropy and scaled relative anisotropy), mean diffusivity, radial diffusivity, eigenvalues and the direction of the first eigenvector are investigated as well.
Original language | English (US) |
---|---|
Pages (from-to) | 777-788 |
Number of pages | 12 |
Journal | Magnetic Resonance Imaging |
Volume | 29 |
Issue number | 6 |
DOIs | |
State | Published - Jul 2011 |
Externally published | Yes |
All Science Journal Classification (ASJC) codes
- Biophysics
- Biomedical Engineering
- Radiology Nuclear Medicine and imaging
Keywords
- Diffusion tensor imaging (DTI)
- MRI
- Noise
- Tensor estimation